甲醛分析方法验证方案
更新时间:2024-06-20 10:20:01 阅读量: 综合文库 文档下载
工作场所空气中甲醛分光光度法验证方案
方案编号:
签名 日期 方案执行日期: 年 月 日 签名: 起草人 审核人 批准人
ZQ ZQ(J)-YZ-2017-2012
目 录
1. 概述 ········································································································ 1 2. 目的 ········································································································ 1 3. 分光光度计操作条件: ················································································ 1 4.检测方法 ··································································································· 1
4.1 试剂 ······················································································································ 1 4.2开机准备 ················································································································ 1
5方法验证 ··································································································· 2
5.1 线形范围 ················································································································ 2 5.2 检出限 ··················································································································· 3 5.3 定量下限 ················································································································ 3 5.4 最低检出浓度 ·········································································································· 4 5.5 精密度 ··················································································································· 4 5.6 准确度 ··················································································································· 4
6.验证结论 ··································································································· 5
附录1:验证所需设备和仪器仪表汇总 ··············································································· 5 附录2:验证原始记录 ···································································································· 5
ZQ
1. 概述
本文件验证了工作场所空气中甲醛的分光光度法本检测方法实施细则参照 GBZ/T 160.54-2004 《工作场所空气有毒物质测定脂肪族醛类化合物》中甲醛的酚试剂分光光度法的定量测定为主要依据,以及在此标准的基础上根据本检测中心实际配置的仪器和实验条件的情况下将标准中的技术要素编写了相关的作业指导书。
为了保证此分析方法的可行性、准确性、可操作性和适用性、用科学的方法进行相应的验证程序特别编写了此验证方案,为今后在实际工作中将起到指导和借鉴的作用。
在验证空气中甲醛的测定方法中,做了方法的线性范围实验、检出限实验、方法精密度实验、准确度实验,以证明该方法适用于测定工作场所空气中甲醛的的浓度。 2. 目的
通过验证工作,确保测定工作场所空气中甲醛浓度的分析方法在广德众康职业卫生检测服务有限公司适用。 3. 分光光度计操作条件:
仪 器 波 长 比色皿 参比溶液
4.检测方法 4.1 试剂
实验用水为蒸馏水
4.1.1酚试剂:称量0.10g酚试剂,加水溶解,倾于100ml具塞量筒中,加水 至刻度。
4.1.2: 硫酸铁铵溶液:称量1.0g硫酸铁铵,用0.72mol/L硫酸溶解,并稀释至100ml。 4.1.3:标准溶液:量取2.8mL含量为36%~38%甲醛溶液,放入1L容量瓶中,加水稀释至刻度。此溶液1ml约相当于1mg甲醛。标定后溶度为0.96mg/ml的贮备液。临用前用水稀释至1.0ug/ml甲醛标准溶液。
1/5
紫外可见分光光度计(T6新世纪) 可调波长范围190~1100nm 645nm 10*10mm 蒸馏水 ZQ
4.2开机准备
4.2.1 检查仪器,准备10mm石英比色皿。
4.2.2 打开紫外可见分光光度计,同时预热分光光度计10min
4.2.3 预热完毕,仪器自检完毕后,准备实验。
4.2.4 在成套的石英比色皿中加入参比溶液,做基线校准与校准能量。 4.2.5 校准完毕后,输入本方法所需要的波长645nm,确认,实验开始。 5.方法验证 5.1 线形范围
在3操作条件下将紫外可见分光光度计调节至最佳测定状态,标准曲线的绘制:取7只具塞比色管,分别加入0.00、0.1、0.25、0.5、1.00、1.50、2.00ml甲醛标准溶液,加水至2.6ml,各加2ml酚试剂溶液,摇匀,于43±1℃水浴中放置10min,期间摇动几次,加入0.4ml硫酸铁铵溶液,摇匀,再放入水浴中加热10min,取出放冷至室温,在645nm 波长下测量吸光度。每个浓度重复测定3 次,以吸光度均值对相应的甲醛的含量(?g)绘制标准曲线。
见表1:
表1:方法验证线性范围数据表
序号 标准含量(μg) 吸光度(A) 相关系数 线形方程 1 0.00 0.031 2 0.1 0.089 3 0.25 0.176 4 0.5 0.307 0.9980 Y=0.5036X+0.046 5 1.00 0.568 6 1.50 0.810 7 2.00 1.036 2/5
ZQ
图1:甲醛的含量与吸光度关系图
5.2 检出限
在3工作条件下将仪器调至检测状态,连续测量10次空白溶液,按公式计算检出限:
CL = 3σ/s
式中:CL — 检出限,?g/mL;
σ — 测量10次空白溶液的浓度标准偏差; s — 方法的灵敏度,即工作曲线斜率,mL/μg。
表2:连续测定10次空白溶液检出线数据
序号 吸光度值 浓度(μg /mL) 序号 吸光度值 浓度(μg /mL) 浓度平均值(mg/L) 标准偏差 1 0.031 0.00 6 0.031 0.00 2 0.031 0.00 7 0.030 0.00 0.00 0.00042 3 0.030 0.00 8 0.031 0.00 灵敏度(mL/μg) 检出限(μg/mL) 4 0.031 0.00 9 0.031 0.00 5 0.031 0.00 10 0.031 0.00 0.5036 0.0025 从表2得到本方法检出限结果:0.0025μg/mL ,低于方法检出限0.04μg/mL,符合检测要求。 5.3 定量下限
3/5
ZQ
LLOQ=10×CL
LLOQ —定量下限,mg/L CL — 检出限,mg/L
因为本方法检出限CL为0.0025mg/L,故定量下限结果:10×0.0025=0.025mg/L
结论:本方法定量下限为0.67mg/L 5.4 最低检出浓度
表3:本方法验证的最低检出浓度测定数据 采样时间 (min) 15 采样流量 (L/min) 02 采样体积 (L) 3 样品体积 (mL) 5 最低检出浓度 (mg/m3) 0.0015 从表3得到本方法最低检出浓度结果:0.0015mg/m 低于方法最低检出浓度
3
0.067mg/m3(以采集3L空气样品计),符合检测要求。
5.5 精密度
在3的工作条件下,对0.25μg的甲醛标准溶液进行6次重复测量,得到吸光度并计算其相对标准偏差。
RSD=σ/?x * 100%
RSD — 相对标准偏差,%; σ — 标准偏差;
?x — 计算结果的算术平均值。
表4:精密度验证结果数据表
样品 测定值A 1 0.176 2 0.176 3 0.177 4 0.176 5 0.176 6 0.176 标准偏差 ?x RSD% 含量μg 0.25 0.25 0.25 0.25 0.25 0.25 0.00041 0.176 0.23 5.6 准确度
5.6.1 质控样品的测定:
用10mL干燥洁净移液管,准确移取浓样10.0mL于100mL容量瓶中,用纯水定容,充分混匀。
4/5
ZQ
质控样品:取6 只具塞比色管,分别加入1.0ml质控溶液,加水至2.6ml,各加2ml酚试剂溶液,摇匀,于43±1℃水浴中放置10min,期间摇动几次,加入0.4ml硫酸铁铵溶液,摇匀,再放入水浴中加热10min,取出放冷至室温,其标准值为0.10ug/mL,不确定度为±0.01 ug/mL.在645nm 波长下测量吸光度。
表5:质控标准样品的测定结果 样品 测定值A 含量μg 浓度 μg/mL 1 2 3 4 5 6 0.574 1.05 0.105 平均值μg/mL 0.105 0.574 0.574 0.574 0.573 0.574 1.05 1.05 1.05 1.04 1.05 标准值 μg/mL 0.10 不确定度 μg/mL 0.005 测定结果 0.105 0.105 0.105 0.104 0.105 符合 结论:质控结果在证书范围内,准确度满意。 6.验证结论
本方法验证结果表明:本法的检出限为0.0025μg/mL;最低检出浓度为0.0015mg/m3( 以采集75L空气样品计)。测定范围为 0.04μg/mL-2.0g/mL;溶液相对标准偏差为 0.0023 ;质控样品检测结果:0.105μg/mL,符合要求。本验证方法适用于验证工作场所空气中甲醛分光光度法。
附录1:验证所需设备和仪器仪表汇总 设备/仪器名称 设备/仪器编号 设备/仪器状态 T6新世纪 正常 证书编号 完成时间 附录2:验证原始记录
————————————————————————————
5/5
正在阅读:
甲醛分析方法验证方案06-20
环境因素识别、更新表103-08
煤矸石综合利用项目可行性研究报告04-03
浅谈国内外绘本研究12-09
驻马店选调生网02-07
17 iPACS-5773测控装置技术说明书V2.01 - 图文03-11
丝光绒面料项目可行性研究报告(发改立项备案+2013年最新案例范文)详细编制方案03-16
北师大版小学数学第八册第一单元测试题08-07
2016年光学薄膜调研及发展前景分析05-02
- 多层物业服务方案
- (审判实务)习惯法与少数民族地区民间纠纷解决问题(孙 潋)
- 人教版新课标六年级下册语文全册教案
- 词语打卡
- photoshop实习报告
- 钢结构设计原理综合测试2
- 2014年期末练习题
- 高中数学中的逆向思维解题方法探讨
- 名师原创 全国通用2014-2015学年高二寒假作业 政治(一)Word版
- 北航《建筑结构检测鉴定与加固》在线作业三
- XX县卫生监督所工程建设项目可行性研究报告
- 小学四年级观察作文经典评语
- 浅谈110KV变电站电气一次设计-程泉焱(1)
- 安全员考试题库
- 国家电网公司变电运维管理规定(试行)
- 义务教育课程标准稿征求意见提纲
- 教学秘书面试技巧
- 钢结构工程施工组织设计
- 水利工程概论论文
- 09届九年级数学第四次模拟试卷
- 甲醛
- 验证
- 方案
- 方法
- 分析
- 初中英语代词专项讲解与训练
- 安徽省地方标准《先张法预应力混凝土管桩基础技术规程》(DB34 50
- 2018-2019年高中教师教学工作总结范文
- 我运动我健康演讲稿
- 教材解析 - 必修2第八课《走进国际社会》
- 石脑油烃类组成
- 高起专 网络教育计算机基础1考试试题及答案
- 管理人破产程序工作文书样式(试行)
- 机电与自动化专业STM32简介Introduction-of-STM32大学毕业论文外
- 银行间债券市场现券交易净额清算业务指南
- 审查指南(2010)5.10
- 公关策划方案范文(试题一)
- 2012学年“温岭市中小学生智能培植与展示系列活动”之 小学语文
- 我的情绪我做主
- 市场调查与预测 重点 知识点 考试题
- 火电建设工程质量监督检查典型大纲(2004年报批稿)
- oracle练习-SQL
- 大学物理3习题湘潭大学
- 毛概社会调查报告 - 图文
- 光学参考答案(改)1