CCT-3320T 电导率表说明书英文版(原CM230)

更新时间:2023-05-15 17:06:01 阅读量: 实用文档 文档下载

说明:文章内容仅供预览,部分内容可能不全。下载后的文档,内容与下面显示的完全一致。下载之前请确认下面内容是否您想要的,是否完整无缺。

ROC

CCT-3320T

Conductivity(TDS)Monitor ®

Manual

1. Specifications Instrument Specifications

Item

Measuring

Ranges Measuring Range(S/cm Electrode Constant(cm-1) 0.5~199.9 1.0~1999

5~9999

Conductivity

Temperature 0.100 1.000 5.000 1.5%(FS) ≤±0.8℃ Accuracy

Tem. Range

4—20mA

Direct

Current

Power In.

Power

Consumption

environmenta

l conditions

Dimension

Install Hole 0~50℃;Compensation base on 25℃ Channel Technical Feature Translate Accuracy 1 Isolate、Reversible、Fully adjustable 1.5%(FS) AC220V±10% 50/60Hz ≤2.5W Temperature:0~50℃;Humidity≤85%RH (48×96×80)mm(H×W×D) (44×92)mm(H×W)

Electrode Parameter

Liquid Tem. (5~50)℃

Screw Size

Liquid

Pressure

Cable

Length 1/2" G 0~ 0.5Mpa 5m OR as customized

2.Backside Wiring Diagram and Description

W—Electrode white line; G/B—Electrode green line;

Y—Electrode yellow line; R—Electrode red line;

I+/I-:4~20mA measuring mode, internal power;

T+/T-:4~20mA transmit mode,internal adjusting

module power;

0V/220V:Alternating current AC 220V Inlet;

Emc grounded protection(earthing); :

NC: No connected(empty connect inside);

3.Measuring Electrode Installation

Bubble or static water can cause electrode measuring errors,

ensure electrode inset enough length. In strict accordance with the below installation:

Correct installation method (1) to ensure that the electrode inset enough length into the measured water.

Correct installation method (2) to ensure that the water flow from the small hole of the electrode

Notice:

1.Electrode should be installed in the low position of the pipe, where velocity stability and not easy to accumulation of air bubbles.

2.Electrode horizontal or vertical installation must inset deep into the water.

3.Measurement signal is a weak electrical signal, its cable should go line independently, avoid power line and control line connection in the same set of cable connector or terminal board.

4.If measuring cable need to be longer, please contract length before delivery.

4. Button Introduction

5.Operating Interface Introduction

M" button for 3 seconds automatically enter the Setup menu

Setup No.Setup Item Function

Label flashing, operate the

Electrode "Position" and "Add" button to input

1

constant electrode constant value, press "Enter" to

save and enter the next parameter settings

Decimal point identification flashing,

Constant decimal "Position" key to select position; Press

2

point "Enter" to save and enter next parameter

settings

measuring Unit (PPM, PPT or S/cm,

3 Units select

mS/cm), and press "Enter" to save and

enter next parameter settings

4 corresponding value; "Enter" to save, 4mA Parallelism"Position" to select decimal point, "Enter"

to save and enter next parameter settings

5 corresponding value; "Enter" to save, 20mAParallelism"Position" to select decimal point, "Enter"

to save and enter next parameter settings

6.Trouble Shooting

When there is measurement data is not correct, unstable faults, perform the following inspection and judgment:

1.To analyze problems caused by instrument or electrode ?

White on the meter terminal line (White) to be removed, display show the measuring is zero, excluding instrument failure, review the

installation of the electrode.

2.Electrode installation and faults

Uninstall the electrode from the pipe, use of known conductivity water to test, such as the correct value for consideration fault cause by the installation.

7.Maintenance

Measuring electrode is precision components decomposition, do not change the conductivity pool shape and size; Do not use strong acid, alkali cleaning, in order to avoid changing electrode constants, affect the measure precision of the instrument.

1.Regularly use neutral cleanser or dilute hydrochloric acid(10%) for short soak, then wash with pure water, in order to keep the conductivity pool surface clean. Measurement cable is special cable, do not change, so as not to introduce significant error.

2.Instrument should be placed in dry environment or control box, avoid water splash or wet leakage meter or measurement error.

河北科瑞达仪器科技股份有限公司