a-pick up tube improper position
adjust pump head to tube x xx mm press blood flow key to confirm air detector not activated press air detector key air leakage check fluid tubes connections attach fluid tubes press select to abort the test attach fluid tubes press select to start disinf b/acetate-pick up tube improper position battery failure change battery or press select bicart emptied press select to confirm
bicart holder not closed close or check bicart holder bicart not attached to holder attach or check bicart blood in priming detector function check stopped
blood leak detector failure clean and press blood leak key blood pump cover not closed close blood pump cover blood pump overload in fch check the pump and press select blood pump overload press blood pump key blood pump stop time expired start blood pump bpm failure turn off bpm
chem dis fill up phase completed place pick up tube in machine chemical fill up not sufficient check supply of disinfectant connect concentrate when time to prepare machine for treatment decalcification required xx hours since last decalcific dialysis flow too low
dialysis fluid not ready for treatment fluid ready in xxx s dialysis fluid ready press bypass key
disconnect venous pressure press venous key to confirm disinfection required xx days since last disinfection extra priming volume achieved
fluid bypass time expired press bypass key fluid path obstruction check fluid tubes and dialyzer fluid tube sensor test failed press select to confirm
fluid tube sensor test attach fluid tubes fluid tube sensor test remove fluid tubes
fluid tubes in safety bypass connect fluid tubes to dialyzer fluid tubes not in safety bypass function check restarted please wait
heater efficiency not sufficient for proper disinfection heparin pump overload check heparin syringe and line heparin set to zero press heparin key to confirm high conductivity not obtained check concentrate high inlet water cond xx ms/cm check inlet water high monitor temperature alarm turn off the monitor impossible time or uf setting adjust time or uf volume incorrect acidic concentrate check contrate incorrect bicarbonate conc check concentrate incorrect concentrate check concentrate incorrect dialysis fluid composition
insufficient inlet pressure please check water supply low battery power machine turns off within one minute negative uf rate check and press uf volume no backfiltration warning press tmp to confirm no bpm values check bpm cuff
no cuff attached attach or check bpm cuff
only blood monitor is running press select to start fluid part dlease wait 30 sec. to bypass dialysis fluid
power failure, battery operation battery operated for xx minutes ready for venous pressure test press venous key to start recovered from power failure press time key to confirm remove fluid tubes press select to abort the test sags test failed
set dialysis flow not reached readjust flow or call service tmp limits too wide press tmp key to adjust
to empty bicart, press select
to end treatment, press isol ufand zero time to obtain total uf volume isol uf must be started to start the air detector test: fill the venous drip chamber too high tmp required increase time or decrease uf volume treatment started no reset made press select if ok, else resetuf uf rate limit reached re-adjust time or uf volume
uf rate lower than minimum set adjust time, uf volume or min uf uf stop time expired press start/stop key
uf volume set to zero press uf volume key to confirm venous low limit below +10 mmhg press venous key to adjust venous pressure limits too wide press venous key to adjust venous pressure test interrupted press select to confirm wrong disinfectant agent press select to restart disinf
A-PICK UP TUBE IMPROPER POSITION A管位置不对
ADJUST PUMP HEAD TO TUBE X XX MM PRESS BLOOD FLOW KEY TO CONFIRM 改动了血泵管径,压血流量键确认
AIR DETECTOR NOT ACTIVATED PRESS AIR DETECTOR KEY 空气监测没启动,压空气监测键
AIR LEAKAGE CHECK FLUID TUBES CONNECTIONS 漏气,检查透析液管路接头
ATTACH FLUID TUBES PRESS SELECT TO ABORT THE TEST 接上透析液管路压箭头终止测试
ATTACH FLUID TUBES PRESS SELECT TO START DISINF 接上透析液管路压箭头开始消毒
B/ACETATE-PICK UP TUBE IMPROPER POSITION B管位置不对
BATTERY FAILURE CHANGE BATTERY OR PRESS SELECT 电池失效,换电池或压箭头键
BICART EMPTIED PRESS SELECT TO CONFIRM 干粉简排空,压键头确认
BICART HOLDER NOT CLOSED CLOSE OR CHECK BICART HOLDER 干粉支架未关好
BICART NOT ATTACHED TO HOLDER ATTACH OR CHECK BICART 干粉未接上支架,请接上
BLOOD IN PRIMING DETECTOR FUNCTION CHECK STOPPED 预冲探测器监测血或预冲探测器赃了
BLOOD LEAK DETECTOR FAILURE CLEAN AND PRESS BLOOD LEAK KEY 漏血测试失败,清洗并压漏血键
BLOOD PUMP COVER NOT CLOSED CLOSE BLOOD PUMP COVER 血泵门未关
BLOOD PUMP OVERLOAD IN FCH CHECK THE PUMP AND PRESS SELECT 自检时血泵卡死,检查血泵压箭头
BLOOD PUMP OVERLOAD PRESS BLOOD PUMP KEY 血泵卡死,压血泵键
BLOOD PUMP STOP TIME EXPIRED START BLOOD PUMP 血泵停止时间过长,启动血泵 BPM FAILURE TURN OFF BPM 血压计坏了,关闭血压计
CHEM DIS FILL UP PHASE COMPLETED PLACE PICK UP TUBE IN MACHINE 消毒液吸满了,将吸液管放入机器内
CHEMICAL FILL UP NOT SUFFICIENT CHECK SUPPLY OF DISINFECTANT 消毒液不对,检测消毒液
CONNECT CONCENTRATE WHEN TIME TO PREPARE MACHINE FOR TREATMENT 请接透析液
DECALCIFICATION REQUIRED XX HOURS SINCE LAST DECALCIFIC 需要脱钙
DIALYSIS FLOW TOO LOW 透析液流量太小
DIALYSIS FLUID NOT READY FOR TREATMENT FLUID READY IN XXXs 透析液未准备好
DIALYSIS FLUID READY PRESS BYPASS KEY 透析液浓度好了,压旁路键
DISCONNECT VENOUS PRESSURE PRESS VENOUS KEY TO CONFIRM 取下静脉压保护器,压静脉压键确认
DISINFECTION REQUIRED XX DAYS SINCE LAST DISINFECTION 需要消毒
EXTRA PRIMING VOLUME ACHIEVED 预冲量到了
FLUID BYPASS TIME EXPIRED PRESS BYPASS KEY 旁路时间太长,压旁路键
FLUID PATH OBSTRUCTION CHECK FLUID TUBES AND DIALYZER 管路中有堵塞,检查管路和透析器
FLUID TUBE SENSOR TEST FAILED PRESS SELECT TO CONFIRM 旁路管位置感应器测试失败,压箭头确认
FLUID TUBE SENSOR TEST ATTACH FLUID TUBES 旁路管位置感应器,接上旁路
FLUID TUBE SENSOR TEST REMOVE FLUID TUBES 旁路管位置感应器,取下旁路
FLUID TUBES IN SAFETY BYPASS CONNECT FLUID TUBES TO DIALYZER 旁路管在机器上,接上透析器
FLUID TUBES NOT IN SAFETY BYPASS 旁路管不在机器上
FUNCTION CHECK RESTARTED PLEASE WAIT 机器自检重新开始
HEATER EFFICIENCY NOT SUFFICIENT FOR PROPER DISINFECTION 热消毒功率不够,检查电压和保险
HEPARIN PUMP OVERLOAD CHECK HEPARIN SYRINGE AND LINE 肝素泵过载,检查针筒到底和管路夹死
HEPARIN SET TO ZERO PRESS HEPARIN KEY TO CONFIRM 肝素设定为零,压甘素键确认
HIGH CONDUCTIVITY NOT OBTAINED CHECK CONCENTRATE 高电导达不到,检查A液
HIGH INLET WATER COND XX ms/cm CHECK INLET WATER 进水电导太高
HIGH MONITOR TEMPERATURE ALARM TURN OFF THE MONITOR 机器高温报警,关机器
IMPOSSIBLE TIME OR UF SETTING ADJUST TIME OR UF VOLUME 时间或脱水设定不对,调节时间或脱水量
INCORRECT ACIDIC CONCENTRATE CHECK CONTRATE A液不对
正在阅读:
血液透析机器报警中英对照04-04
权力的定义、起源、性质、利弊04-29
中国地质大学高等数学A1期末试卷A08-29
三环与四环立交桥监理细则07-03
粉体材料基础实验指导书 - 图文06-03
平面计划培训费用大约是几多钱(整理规划师学习班多少钱)03-30
2011部分高校MTI参考书目11-03
ORACLE表备份恢复测试用例10-28
- 多层物业服务方案
- (审判实务)习惯法与少数民族地区民间纠纷解决问题(孙 潋)
- 人教版新课标六年级下册语文全册教案
- 词语打卡
- photoshop实习报告
- 钢结构设计原理综合测试2
- 2014年期末练习题
- 高中数学中的逆向思维解题方法探讨
- 名师原创 全国通用2014-2015学年高二寒假作业 政治(一)Word版
- 北航《建筑结构检测鉴定与加固》在线作业三
- XX县卫生监督所工程建设项目可行性研究报告
- 小学四年级观察作文经典评语
- 浅谈110KV变电站电气一次设计-程泉焱(1)
- 安全员考试题库
- 国家电网公司变电运维管理规定(试行)
- 义务教育课程标准稿征求意见提纲
- 教学秘书面试技巧
- 钢结构工程施工组织设计
- 水利工程概论论文
- 09届九年级数学第四次模拟试卷
- 血液透析
- 中英
- 对照
- 报警
- 机器
- 北邮大学英语3(新) 阶段作业一
- 各国的节日,代表人物,标志物
- 2016年山东公务员考试行测真题及参考答案
- 上海高考英语4月试卷
- 2014年广东大学生数学建模竞赛初评结果-华南师范大学 - 图文
- 河南省卫生厅关于印发《河南省护理事业发展规划纲要(2011-2015
- 生物安全操作规程和技术规范
- 2005年4月新兴县高考备考研讨会发言稿
- 《传承好家训弘扬好家风》亲子共读经典家风家训
- 计算机软件毕业论文
- i9s win8刷机方法
- 2018年中国香肠现状调研及市场前景预测(目录) - 图文
- 战略成本管理案例分析—以新希望集团为例
- 第3版新视野大学英语4读写教程课本练习答案(全) - 图文
- 五年级奥数 计数综合 容斥原理(ABC级).学生版
- 消防安全技术综合能力
- 新生儿科题目(1)
- 公司党支部二季度员工思想动态分析
- 继续教育公需科目必学科目答案
- 学习《慈善法》的几点认识