NonClinical Dose Formulation Analysis Method Validation and Sample Analysis
更新时间:2023-05-29 05:23:01 阅读量: 实用文档 文档下载
毒理学实验供试品分析指导原则白皮书
White Paper
NonClinical Dose Formulation Analysis Method Validation and Sample Analysis
Monica Lee Whitmire,1,8Peter Bryan,2Teresa R.Henry,3John Holbrook,Paul Lehmann,4Thomas Mollitor,Susan Ohorodnik,5David Reed,6and Holly D.Wietgrefe 7
Received 30April 2010;accepted 6July 2010;published online 14August 2010
Abstract.Nonclinical dose formulation analysis methods are used to con ?rm test article concentration and homogeneity in formulations and determine formulation stability in support of regulated nonclinical studies.There is currently no regulatory guidance for nonclinical dose formulation analysis method validation or sample analysis.Regulatory guidance for the validation of analytical procedures has been developed for drug product/formulation testing;however,veri ?cation of the formulation concentrations falls under the framework of GLP regulations (not GMP).The only current related regulatory guidance is the bioanalytical guidance for method validation.The fundamental parameters for bioanalysis and formulation analysis validations that overlap include:recovery,accuracy,precision,speci ?city,selectivity,carryover,sensitivity,and stability.Divergence in bioanalytical and drug product validations typically center around the acceptance criteria used.As the dose formulation samples are not true “unknowns ”,the concept of quality control samples that cover the entire range of the standard curve serving as the indication for the con ?dence in the data generated from the “unknown ”study samples may not always be necessary.Also,the standard bioanalytical acceptance criteria may not be directly applicable,especially when the determined concentration does not match the target concentration.This paper attempts to reconcile the different practices being performed in the community and to provide recommendations of best practices and proposed acceptance criteria for nonclinical dose formulation method validation and sample analysis.
KEY WORDS:acceptance criteria;formulation method validation;formulation sample analysis;nonclinical dose formulation analysis;test article concentration and homogeneity.
INTRODUCTION
Nonclinical dose formulation analysis (NCDFA)is required in all regulated studies used to assess the safety of drugs during the development process.The primary purpose of nonclinical studies is to establish safety margins which can then be extrapolated to clinical studies.Therefore,NCDFA is required in all regulated studies to verify the documented test article concentrations in formulations used to calculate these safety margins (1,2).These analytical methods are used to assess the concentration of test article in nonclinical formu-lation,formulation homogeneity and formulation stability in support of regulated nonclinical studies (for example:safety,toxicokinetic,and pharmacokinetic studies).
The in-life phase and therefore the dose formulation analysis phase of regulated nonclinical studies are typically conducted in compliance with one or more of the following:
(1)the Food and Drug Administration (FDA)Good Labo-ratory Practice Regulations (GLP)as set forth in Title 21of the US Code of Federal Regulations,Part 58(1,2);(2)the Organisation for Economic Co-operation and Development (OECD)Principles of Good Laboratory Practice,[ENV/MC/CHEM(98)17](3);(3)the Japanese Ministry of Health,Labor and Welfare (MHLW)No.21,March 26,1997(4);and (4)the FDA Guidance for Industry:Bioanalytical Method Validation,May 2001(5).It should be mentioned that the ?nal guidance listed above,as the name suggests,is speci ?cally for bioanalytical method validation,not dose formulation analysis validation.The fundamental parameters for bioanalytical and NCDFA validations that overlap include:recovery,accuracy,precision,selectivity (speci ?city),carryover,sensitivity and stability (6–9).Divergence in bioanalytical and NCDFA validations typically center around the acceptance criteria used.As the dose formulation samples are not true “unknowns ”,the bioanalytical concept of quality control samples that cover the entire range of the standard curve serving as the indication for the con ?dence of the data generated from the “unknown ”study samples may not always be necessary.Also,the standard bioanalytical acceptance criteria may not be directly applicable,especially when the determined concentration does not match the target concentration.
Since none of the aforementioned guidance or the numerous guidelines that have been written for the validation
1MPI Research,54943North Main Street,Mattawan,Michigan,USA.2
Celgene,Summit,New Jersey,USA.3
Pharmavite,Mission Hills,California,USA.4
Sano ?Aventis,Bridgewater,New Jersey,USA.5
GTx,Inc.,Memphis,Tennessee,USA.6
Merck,West Point,PA,USA.7
Advion,Ithaca,New York,USA.8
To whom correspondence should be addressed.(e-mail:monica.whitmire@)
The AAPS Journal,Vol.12,No.4,December 2010(#2010)DOI:10.1208/s12248-010-9226-9
1550-7416/10/0400-0628/0#2010The Author(s).This article is published with open access at
628
毒理学实验供试品分析指导原则白皮书
of analytical procedures(10–16)speci?cally address the analysis of nonclinical dose formulations,the members of the NonClinical Dose Formulation Analysis Focus Group of the American Association of Pharmaceutical Scientists (AAPS)met to discuss issues associated with NCDFA of small molecules,in June and November of2009.The attend-ees represented a cross-section of the focus group from contract research organizations,pharmaceutical companies and biotech companies with analytical,bioanalytical,toxicology,and formulation experience represented.The intent of these meetings was to gain consensus on nonclinical dose formulation analysis criteria for validation and sample analysis.The resulting consensus is the framework of this white paper.
GOALS AND OBJECTIVES
The purpose of this AAPS NCDFA Focus Group-sponsored white paper is to outline the general requirements for validating analytical methods and performing routine sample analysis in support of nonclinical drug formulation assessments.This paper focuses on method validation and sample analysis;however,the intent is not to minimize the importance of performing good science when developing an analytical method.
The analytical method should be suitable for the intended use/purpose and should generate reliable results that are free from signi?cant bias.These analytical methods will be utilized to provide routine GLP support of con?rma-tion of analyte(test article)dosage concentration,homoge-neity assessment and stability testing.These analytical methods are not primarily intended to provide quanti?cation of minor impurities for use in stability-indicating methods. Analyte and Vehicle
Analyte.An active pharmaceutical ingredient(API)is known as a test article when dosed and as an analyte during method validation and sample analysis.The analyte should be characterized with established purity,storage conditions,and retest or expiration date ideally documented in a certi?cate of analysis(or purity statement).
Vehicle.A vehicle(also known as excipients)is the material(s)used to deliver the test article to a biological system. Examples of vehicles include0.5%methylcellulose,saline, water,feed,etc.Documentation of the vehicle components of the test article system is necessary during method validation. PROPOSED METHOD VALIDATION APPROACH The general approach and criteria for validating methods for dose formulation analysis are instrument-independent. Currently,the majority of formulation analysis methods utilize high-performance liquid chromatography-ultraviolet (HPLC-UV),LC-mass spectrometry(MS)/MS,or GC-FID instrumentation.The suggested criteria provided in the remainder of this document are speci?c to HPLC-UV methodologies,but the underlying principles may be applied to other suitable analytical techniques.NCDFA validation experiments should encompass the anticipated dosage con-centration range and include accuracy,precision,speci?city, standard curve linearity and analytical solution stability(10–19).In addition,formulation stability and homogeneity should be established for the conditions a sample is expected to encounter during sample collection,storage,analysis,and dosing.Each validation experiment is discussed in detail. Before initiating a method validation,a validation protocol, validation plan or appropriate SOP(s)to support the method validation experimentation should be written and approved. The validation protocol should include the acceptance criteria used to pass/fail individual runs and assess other validation experimental results.The method is considered valid if these acceptance criteria are achieved.
An analytical method is often used to support many different studies after it is validated.Therefore the method development phase must be rigorous enough to ensure the ?nal method will be suitable for its intended purpose.The analytical method should cover the entire dose concentration range for all studies that it will be used to support.If a method development study exhibits highly variable results (e.g.,high%RSD)or lower than desired percentage recovery values,various parameters should be assessed prior to proceeding to validation including but not limited to the following:(1)container composition;(2)protection from light;(3)temperature;(4)?lter bias;(5)dosing apparatus, and(6)stability.Assessment of mini-dose formulations or spiked additions of the analyte and vehicle components should be conducted in the method development phase.A typical item that is often overlooked is that low-dose concentrations are made using high amounts of vehicle components and exactly the opposite for high dose concen-trations.Therefore,pH and speci?city for the entire dosing range need to be considered during method development. Solubility is also a critical attribute and depends on the formulation vehicle properties.Most laboratories rely on visual observations for solubility measurements;however, instrumentation exists for more accurate evaluations.Although calculation of correction factors may vary from one compound to the next,correction factors used during validation should be representative of those planned for use in the preparation of study samples.Further de?nition of critical method develop-ment parameters are outside of the scope of this paper. Types of Validations
For all analytical method validations,acceptance criteria must be de?ned prior to the initiation of the validation(10–19). The analytical methods discussed herein are not intended to provide quanti?cation of minor impurities for use in stability-indicating methods.The analytical method should be suitable for the intended purpose.It is conceivable to have a rapid analytical method for test article only,which would not be suitable as a stability indicating assay where degradants are separated from the test article or a more elaborate method which may have the additional advantage of identifying degradants.Both types of assays are acceptable and would require the appropriate level of validation for their intended purposes.
Full V alidation.A full validation is conducted for assay methods used for chronic toxicity studies(>3months in
629
Nonclinical Dose Formulation Analysis Method Validation
毒理学实验供试品分析指导原则白皮书
duration)and encompasses all elements of the validation set forth in this paper.A full validation includes multiple sets of accuracy and precision data.
Early Phase Validation.An early phase validation is conducted for assay methods for acute toxicity studies (≤3months).Early phase validation testing may include a single validation run due to time constraints and limited availability of API.An early phase validation will assess system suitability,linearity and range,accuracy,speci?city,selectivity, and carryover in one analytical run.Precision data obtained may be limited as replicates and multiple runs may not be performed.
Partial Validation.A partial validation should be con-ducted for a validated method when there is a signi?cant change in the method.These changes may include but are not limited to: vehicle composition,dose formulation concentration range, analytical concentration range,chromatographic conditions, detector type,sample-processing procedure,or software.Partial validations are sometimes known as method quali?cations and require a minimum of one set of accuracy and precision data.
Transfer Validation.A transfer validation should be conducted when transferring a fully validated method from one laboratory to another which utilizes the same method, same vehicle,same validation range,and same prede?ned acceptance criteria.Transfer validations require a minimum of one set of accuracy and precision data.
System Suitability Test
Scienti?cally quali?ed and properly maintained instru-ments should be used for implementation of analytical methods in routine dose formulation analysis.Performance of system suitability ensures that the system is operating properly at the time of analysis.System suitability tests are more appropriately used for chromatographic methods to ensure that the system is suf?ciently sensitive,speci?c and reproducible for the current analytical run.Examples of typical system suitability test(SST)factors are injection precision(retention time and peak area),theoretical plates (N),tailing factor(T),capacity factor(k′),and resolution. Stock Standard Comparison
The accuracy of standard preparation should be demon-strated by comparing the response of two separately weighed stock solutions which have been diluted to a single concen-tration within the linear range of the method.Only stocks which compare within5%difference should be used to prepare subsequent standards or quality control samples. Performance Check Standards
The system suitability check samples or quality control samples prepared in vehicle should be injected periodically throughout every chromatographic run to assess consistent analytical performance and serve as a performance check.The typical performance measure for performance check standards is accuracy.
Calibration Curves
A calibration curve is assessed as a function of the detector response at known concentrations of analyte:the assay's linear range(using a particular regression formula)is the lowest to the highest diluted concentration.Standards are usually prepared as simple solutions of analyte in diluent.If signi?cant bias exists,standards should be prepared in vehicle. Linearity and Range
The formulation range is the concentration of the API in the on-study formulation.The analytical range is the vali-dated linear range of the analytical method within which the dose formulation samples will be diluted.Linearity should be demonstrated over the entire analytical range.The linearity data may be used to support a single-point or multi-point calibration curve provided the appropriate criteria are met.
Single-point or multi-point calibration curves may be used if acceptable linearity(3or more concentrations)is demonstrated during validation with acceptable coef?cient of determination values(>0.99).If the y-intercept is signi?cantly different from zero,then the standard curve should be prepared with the vehicle. Recovery/Accuracy/Precision
Accuracy(recovery):Accuracy is the closeness of agree-ment between the average of replicate test results and the nominal value,measured in terms of percentage of recovery, relative error,or deviation from theoretical.
One approach to validate recovery or accuracy is based on small scale preparations of analyte in vehicle which mimic formulation preparations.These preparations should be made in vehicle at a minimum of the low and high concentrations with respect to the anticipated dosing range.The preparations should then be diluted to a target concentration within the anticipated analytical range.Multiple preparations or dilu-tions from a stock may be used to assess precision.
Another approach to validate recovery and accuracy is based on spike preparations of analyte and vehicle in the diluent.This approach is helpful when validating the low end of the formulation range where weighing requirements cannot be met or when homogeneity of mimic formulations is not likely.
Intra-run and Inter-run Accuracy
Intra-run accuracy should be established during validation at a minimum.Inter-run accuracy should be performed in situations where there is a complex vehicle(e.g.,suspensions, solids and feeds).Inter-run accuracy should be measured using a minimum of three determinations per concentration.The analytical method should result in accuracy values of100±10%recovery for solutions,100±15%recovery for suspensions, and100±20%recovery for solids where recovery is measured by dividing the found concentration by the nominal value. Values outside of this range may be acceptable if recovery is consistent across the concentration range.
630Whitmire et al.
毒理学实验供试品分析指导原则白皮书
Intra-run and Inter-run Precision
Intra-run precision should be established during valida-tion at a minimum.Precision is the closeness of agreement (degree of scatter)between replicate independent test results, measured in terms of relative standard deviation(%RSD)or coef?cient of variation(%CV).Precision should be measured using a minimum of three preparations per concentration. The analytical method should result in precision values of≤5%RSD for solutions,≤10%RSD for suspensions,and≤20%RSD for solids
Specificity/Selectivity/Carryover
Speci?city or selectivity is the degree to which a method can quantify the analyte accurately in the presence of inter-ferents (e.g.,vehicle components,impurities and degradants).For chromatographic procedures,representative chromatograms should be used to demonstrate speci?city(e.g.,diluent blank, vehicle blank,analyte in diluent,and analyte in vehicle).
A method should be shown to be speci?c and selective in that it should be able to quantify the analyte accurately in the presence of potential interferences from degradants,the vehicle and diluent.There should be no signi?cant peaks at the retention time of the analyte in diluent or vehicle blanks. Any carryover should be minimized in order to increase the ability to detect analyte in the vehicle control samples. Methods utilizing single-point calibration curves should have carryover no greater than1%of the target standard concen-tration.Methods involving multi-point calibrations should have a carryover response of no more than20%of the limit of quanti?cation(LOQ).
Sensitivity
The LOQ of an assay should be de?ned as the lowest concentration at which an assay is validated.All values that are below the LOQ should be reported as<LOQ and not0to allow the correct interpretation of the results.If samples are diluted as part of the analysis,the dilution should be incorporated into the reporting of the LOQ.The LOQ should have a signal to noise ratio of≥10.
Stability Recommendations:Preprocessed,Postprocessed, Storage,Stock Solution
Stability of the analyte during storage of the test article prior to use,during use and throughout sample preparation and analysis should be demonstrated.Additionally,stock solution stability should be assessed when solutions are stored at room temperature,refrigerated or frozen for a relevant period of time.Stability should be compared with the nominal concentrations of freshly prepared standards.Acceptance criteria are typically based on the formulation vehicle compo-sition:for example,solutions¼100Æ10%recovery with 10%RSD,suspensions¼100Æ15%recovery with10% RSD and solids¼100Æ20%recovery with15%RSD.
Preprocessed Stability.Stability should be generated so as to cover all likely temperatures and times that the samples will be exposed as part of the in-life administration or sample analysis portions of a study.Generally1to48h of room temperature or refrigerated storage is suf?cient.
Postprocessed Stability.Stability of the processed sam-ples should be established to con?rm the integrity of the samples after storage(for example,within the autosampler) in the case where a reinjection is required due to instrument malfunctions or power outages.Typically ambient and/or refrigerated conditions are investigated for the time it takes to perform the second chromatographic run(2to3days).
Storage Stability.Stability used to determine formulation storage conditions and expiration dates.Storage stability is used to determine how long formulations may be used for dosing as well as to support storage prior to sample analysis. Stability must be established to cover the time from sample preparation to sample analysis.
Freeze/Thaw Stability.Based on the anticipated storage conditions and sample analysis procedure,it may be neces-sary to evaluate freeze/thaw stability.If evaluated,stability should be generated so as to cover the number of times/ conditions under which a sample is likely to be frozen and thawed during shipment and analysis.Although most samples only experience one freeze/thaw cycle,reanalysis,or inadver-tent thawing during transport may occur.In such cases, stability should be checked over multiple cycles of freezing and thawing at the relevant conditions.
Stock Solution Stability.The stability of stock solutions of the analyte should be evaluated at room temperature for at least6h.If the stock solutions are refrigerated or frozen for a relevant period,the stability should be evaluated.After completion of the desired storage time,the stability should be assessed by comparing the instrument response of the test solution with that of a freshly prepared solution.Only stocks which compare within5%difference should be used to prepare subsequent standards or quality control samples. Effective and Efficient Documentation—Validation Summary Report Minimal Contents
All method validation data should be summarized in a validation summary or report.The validated method should be followed as written during sample analysis. FORMULATION SAMPLE ANALYSIS REQUIREMENTS
GLPs require testing of the formulations to assure accurate concentration,uniform homogeneity and stability at the intended storage conditions for the duration of use and storage.All formulation samples should be assessed within the validated storage time frame following the analytical method as written.Acceptance criteria are typically based on the formulation vehicle composition:for example,solutions¼100Æ10%recovery with10%RSD,suspensions¼100Æ15%recovery with10%RSD and solids¼100Æ20% recovery with15%RSD.
631
Nonclinical Dose Formulation Analysis Method Validation
毒理学实验供试品分析指导原则白皮书
Sample Analysis Procedures
The following components should be included as part of a study sample analysis run:
System Suitability Test
The“system suitability”should be demonstrated before initiating a study sample batch analysis.The system suitability tests performance and criteria should have been demonstrated during validation and incorporated into the analytical method. Typically,standards or quality control samples are used to con?rm acceptable sensitivity and reproducible response while blanks are used to con?rm there is no interference.Examples of typical SST factors are injection precision(retention time and peak area),theoretical plates(N),tailing factor(T),capacity factor(k′),and resolution.
Calibration Curve
A single-point calibration curve may be used if all formulations are diluted to the same concentration as the single-point calibrator and the curve has been shown to be linear.A multi-point calibration curve should have multiple concentration levels and show suitable response over the analytical range,demonstrating an acceptable coef?cient of determination value.
Stock Standard Comparison
The accuracy of weighing should be demonstrated by comparing the response of two separately weighed stock solutions which have been diluted to a single concentration within the linear range of the method.Only stocks which compare within5%difference should be used to prepare subsequent standards or quality control samples. Performance Check Standards
The performance of the method should be assessed over the course of each analytical run.For this purpose,a single dilution of a stock solution or of a quality control solution in vehicle may be made to the target concentration and injected multiple times during the run.The measured concentration should be within±5%of the target concentration.Any samples not bracketed by acceptable performance check standards or calibration curve standards should be considered nonreportable and reanalyzed.
Types of Dose Formulation Study Samples
Concentration Analysis
Concentration assessments should be performed for every dosage concentration including control/vehicle sam-ples,at a minimum for the?rst and last test batches.More frequent assessments during the course of the study may be required by the SOP/protocol.Long-term(3months or longer)studies typically assess concentration throughout the course of the study at prede?ned intervals(for example,once a month or beginning,middle,and end).If replicate samples are analyzed,the individual values and the average%recovery value are reported as well as the% RSD.If fewer than three samples are analyzed,then a% RSD will not be reported.
Homogeneity Analysis
Homogeneity assessments are required for all formula-tions at study initiation with the exception of“true solutions.”Homogeneity assessments are usually performed for the?rst test batch low and high dosage form concentrations and whenever the batch size changes signi?cantly(for example, homogeneity is usually repeated when there is a20–50% change in batch size).Homogeneity assessment may also be conducted as part of a validation to con?rm whether a“true solution”or suspension has been prepared.Homogeneity is usually performed by assessing replicate samples from the top,middle and bottom strata of the dosage form preparation vessel.In addition to the measured concentration of each sample,the average and%RSD of all aliquots analyzed from a single preparation should be reported.
Resuspension homogeneity should be performed when a formulation is prepared,stored,and used on a daily basis over a period of time.Over time,a formulation may settle without mixing or precipitate if stored at a temperature less than the conditions for preparation.Resuspension homogeneity is usually performed by assessing replicate samples from the top,middle and bottom strata of the dosage form storage vessel after a de?ned period of time and storage conditions.
Stability Analysis
Stability assessments are usually performed during the assay validation or during the toxicology studies by collecting samples from at least the lowest and highest dosage concentrations.Replicate samples are stored at the conditions to be used during the in-life phase of the study to cover the time from preparation to the time of?nal dose analysis.
STUDY SAMPLE COLLECTION
Dose formulation study samples may be solutions, suspensions(for example,microemulsions)or solids.The method of collection of study samples can be critical to accurate analysis.For example,an analyte in a micro-emulsion may fall out of suspension before analysis is performed,but is still chemically stable in the vehicle.In such a situation,taking a sub-aliquot of the study sample may not provide an accurate result.For these types of formulations,special collection procedures should be provided within the study protocol.Exact aliquots where the entire sample is analyzed should be used for suspen-sions.It is also recommended that replicate study samples be collected to ensure suf?cient sample volume for out of speci?cation investigations and repeat analyses for failed runs.If the samples are to be shipped,it is recommended that the back-up samples be shipped separately.
632Whitmire et al.
毒理学实验供试品分析指导原则白皮书
OUT OF ACCEPTANCE/SPECIFICATION INVESTIGATIONS
All study sample results that do not meet acceptance criteria/speci?cation criteria should be investigated.The procedures for an out of acceptance investigation should be clearly de?ned within a SOP,study protocol or plan.The FDA guidance for Industry“Investigating Out-of-Speci?-cation(OOS)Test Results for Pharmaceutical Production”
(20)has been commonly used as guidance for NCDFA OOS.
SUMMARY
NCDFA is an important part of all preclinical studies. Methods used for sample analysis in GLP studies should be validated for accuracy,precision,selectivity and sensitivity in compliance with existing FDA guidelines.Formulations should be assessed for stability and homogeneity before or during GLP studies to ensure sample integrity and reprodu-cibility of results.Close adherence to the recommendations of this paper should be used to insure that robust,reliable and appropriate methods are used for formulation analysis. ACKNOWLEDGEMENTS
The authors would like to thank the following:John Buckholz, B.S.,Brian Burback Ph.D.,Binh Do,Dannette Doolittle,Ph.D.,Rodrigo Laureano,M.S.,Josef Ludwig,B.S., MBA,Emily Mainstone,B.S.,Stacy Marden,Ph.D.,Tony Nakhla, M.S.,Shaundel Percey,B.S.,MT(ASCP),Y uan-Shek Chen, Ph.D.,Phillip J.Wang,Ph.D,MBA,and Laixin Wang,Ph.D. Open Access This article is distributed under the terms of the Creative Commons Attribution Noncommercial License which permits any noncommercial use,distribution,and reproduction in any medium,provided the original author(s)and source are credited.
REFERENCES
1.FDA Nonclinical Laboratory Studies Good Laboratory Practice
Regulations.Available at:http://www.fda.gov/ora/compliance_ref/ bimo/glp/78fr-glp?nalrule.pdf
2.FDA21CFR Part58Good Laboratory Practice Regulations
Final Rule.Available at:http://www.fda.gov/ora/compliance_ref/ bimo/glp/87fr-glpamendment.pdf
anisation for Economic Co-operation and Development
(OECD)Principles of Good Laboratory Practice,[ENV/MC/ CHEM(98)17],26Jan1998.
4.Japanese Ministry of Health,Labor and Welfare(MHLW)No.
21,March26,1997
5.FDA.Guidance for Industry:Bioanalytical Method Validation.
In:U.S.Department of Health and Human Services,Food and Drug Administration,Center for Drug Evaluation and Research, May2001.
6.Shah V,Midha K,Dighe S,McGilveray I,Skelly J,Yacobi A,et
al.Analytical methods validation:bioavailability,bioequivalence and pharmacokinetic studies.Conference report.Eur J Drug Metab Pharmacokinet.1991;16(4):249–55.
7.Shah V,Midha K,Findlay J,Hill H,Hulse J,McGilveray I,et al.
Bioanalytical method validation—a revisit with a decade of progress.Pharm Res.2000;17(12):1551–7.
8.Viswanathan C,Bansal S,Booth B,DeStefano A,Rose M,
Sailstad J,et al.Workshop/conference report—quantitative bioanalytical methods validation and implementation:best practices for chromatographic and ligand binding assays.Pharm Res.2007;24(10):1962–73.Epub2007Apr26.
9.Shah V.The history of bioanalytical method validation and
regulation:evolution of a guidance document on bioanalytical methods validation.The AAPS Journal.2007;9(1):E43–7.Article5.
P32-NF27(United States Pharmacopeial Convention,
Rockville,MD,2009),pp.3988–91.
11.FDA,“Analytical Procedures and Methods Validation:Chem-
istry,Manufacturing and Controls.”Federal Register(Notices) 65(160),52,776-52,777(August2000).
12.Q2A“Text on validation of analytical procedures.”International
Conference on Harmonization.Fed Regist.1995;60:11260 13.Q2B“validation of analytical procedures:methodology.”Inter-
national Conference on Harmonization.Federal Register.1996;
60:11260.
14.Q2(R1)“Validation of analytical procedures:text and method-
ology.”International Conference on Harmonization.Federal Register.1997;62:27463–27467.
15.“International Conference on Harmonization;Draft Guidance
on Speci?cations:Test Procedures and Acceptance Criteria for New Drug Substances and Products:Chemical Substances.”
Federal Register(Notices)2000;65(251):83041–63.
16.“Harmonized guidelines for single-laboratory validation of
methods of analysis.”International Union of Pure and Applied Chemistry(IUPAC)Technical Report.Pure Appl Chem.2002;
74(5):835–55.
17.FDA CDER Guidance for Industry Analytical Procedures and
Methods Validation.Available at:http://www.fda.gov/cder/ guidance/2396dft.pdf
18.FDA Guidance for Industry Good Laboratory Practices Ques-
tions and Answers revised July2007.Available at:http://www.
fda.gov/ora/compliance_ref/bimo/glp/81GLP-qanda.pdf
19.FDA CDER Reviewer Guidance Validation of Chromatographic
Methods.Available at:http://www.fda.gov/cder/guidance/ cmc3_rev.pdf
20.FDA.Guidance for Industry:Investigating Out-of-Speci?cation
(OOS)Test Results for Pharmaceutical Production.In:U.S.Depart-ment of Health and Human Services,Food and Drug Administration, Center for Drug Evaluation and Research,October2006. Definitions
Accuracy Calculations:
%Recovery¼found concentration
ðÞ=nominal concentration
ðÞÂ100 %Relative Error%Deviation
ðÞ¼found concentration
ðÞÀ
½
nominal concentration
ðÞ =nominal concentration
ðÞÂ100 Feed:Feeds or chows are blended from various raw materials and additives and are formulated according to the speci?c require-ments of the target animal.They may be meals,pellets or crumbles.
Intra-run:within one day or analytical run or analytical sequence
Inter-run:between days or analytical runs or analytical sequences
Nominal concentration:The theoretical concentration,cor-rected for salt form and purity(as applicable),of a formulation based on the amount of analyte weighed per total volume of analyte plus vehicle.
Precision Calculations:
%Relative Standard Deviation%Coefficient of Variation
ðÞ¼SDÂ100=Mean
Note:%Relative Standard Deviation is appropriate only for data sets containing3or more points.
Quality Control Sample:A solution,suspension or solid containing test article in formulation vehicle designed to mimic actual dosage formulations.
Solution:A solution is a homogeneous mixture composed of
633
Nonclinical Dose Formulation Analysis Method Validation
毒理学实验供试品分析指导原则白皮书
two or more substances dissolved in a solvent.
Solid:A solid object does not?ow to take on the shape of its container,nor does it expand to?ll the entire volume available. Examples are powders,powders in capsules and tablets.
Suspension:A suspension is a heterogeneous mixture in which solute-like particles may settle out of solvent-like phase some time after their introduction.
True Solution:A“true solution”is a solution in which the analyte is completely dissolved in the liquid phase.If a true solution is?ltered,then the?ltrate and the retentate generate the same concentration value.If a true solution is centrifuged,then no particles are observed.If a true solution is analyzed utilizing a solubility scanner,then the particle size maintains a horizontal axis across the solubility range.
634Whitmire et al.
正在阅读:
NonClinical Dose Formulation Analysis Method Validation and Sample Analysis05-29
反贿赂承诺书03-31
xx社区志愿者服务情况汇报05-13
无行贿承诺书01-05
小机解体报告 - 图文12-16
五一促销广告语02-17
物业管理面临挑战与物业管理创新06-25
《精细化工工艺学》思考题与习题10-23
数字图像处理试题11-13
中考英语复习临考前单句改错100题09-15
- 1Localized Components Analysis
- 2Steps towards an integrated data analysis Basic concepts and Bayesian analysis of Thomson s
- 3A critical discourse analysis of racial
- 4A critical discourse analysis of racial
- 5Opinion mining and sentiment analysis
- 6Numerical Analysis of the Detection Performance
- 7Chapter 4 Time Series Analysis
- 84-Financial Statement Analysis
- 9Analysis of Rhetorical Features in Advertising English
- 10Industry Analysis - Management Consulting in China
- 教学能力大赛决赛获奖-教学实施报告-(完整图文版)
- 互联网+数据中心行业分析报告
- 2017上海杨浦区高三一模数学试题及答案
- 招商部差旅接待管理制度(4-25)
- 学生游玩安全注意事项
- 学生信息管理系统(文档模板供参考)
- 叉车门架有限元分析及系统设计
- 2014帮助残疾人志愿者服务情况记录
- 叶绿体中色素的提取和分离实验
- 中国食物成分表2020年最新权威完整改进版
- 推动国土资源领域生态文明建设
- 给水管道冲洗和消毒记录
- 计算机软件专业自我评价
- 高中数学必修1-5知识点归纳
- 2018-2022年中国第五代移动通信技术(5G)产业深度分析及发展前景研究报告发展趋势(目录)
- 生产车间巡查制度
- 2018版中国光热发电行业深度研究报告目录
- (通用)2019年中考数学总复习 第一章 第四节 数的开方与二次根式课件
- 2017_2018学年高中语文第二单元第4课说数课件粤教版
- 上市新药Lumateperone(卢美哌隆)合成检索总结报告
- Analysis
- NonClinical
- Formulation
- Validation
- Method
- Sample
- Dose
- 福建省紫金山大型铜金矿床的发现与研究
- 十一自驾旅游路书
- 最完美的漳州介绍
- 2010海南省会计从业资格理论考试试题及答案
- 参加笔试人员名单xls - 欢迎进入洛阳银行
- 清华大学大气污染控制工程试题3
- 2014最新更新规范图集目录无超连接板-水
- 2016-2021年中国连翘败毒丸行业市场调研及战略规划投资预测报告
- 钢结构大棚建设施工合同书
- Goodix Oncell Layout 设计规范(20140217)
- 2014年黑龙江公务员考试行测判断推理模拟试题(16)
- 临床用血申请与分级管理制度1
- 外研版九年级英语练习题及答案全套
- 4.影响建筑色彩文化的几个因素
- 电涡流式数字转速表
- 金属材料机械性能
- 三年级数学下册除数是一位数的除法单元测试题(二)
- 新高中数学《集合》专项测试 (302)
- 高一下学期班主任工作计划.
- 教职工代表大会会议记录